In silico intravascular optical coherence tomography (IVOCT) for quality assured imaging with reduced intervention

  • Haoyu Zhuo,
  • Xianchen Huang,
  • Jie Xiao

摘要

In the clinical application of intravascular optical coherence tomography (IVOCT), it is necessary to flush opaque blood during image acquisition. However, there are no specific standards for how to perform low-dose but effective flushing. In this study, computational fluid dynamics (CFD) and optical models were integrated to numerically simulate the complete process of IVOCT, which includes blood flushing with normal saline followed by image acquisition. Moreover, an intermittent injection scheme was proposed, and its advantages over the conventionally adopted scheme of continuous injection were verified. The results show that intermittent injection can significantly reduce the dosage of normal saline (reduced by 44.4%) with only a slight sacrifice of image quality (reduced by 8.7%, but still acceptable). The developed model and key findings in this work can help surgeons practice optimized IVOCT operations and potentially lead to improved designs of the IVOCT equipment.