Modified streptavidin–biotin based lateral flow test strip for rapid detection of SARS-CoV-2 S1 antigen in saliva samples
摘要
Compared to other infectious diseases, for which LFT development can take years, SARS-CoV-2 antigen LFTS were developed and deployed within months. LFTS for antigen detection were adopted on an unprecedented scale during the COVID-19 pandemic, but many of them lack the sensitivity especially for samples with low viral load. In our previous work, we developed an enhanced signal strip for detection of SARS CoV-2 SI antigens in saliva. Here we introduce some modification to improve the sensitivity, and specificity, and to lower the cost of the strip, by using biotin streptavidin (BS) system. In the modified BS strip, gold-streptavidin and biotinylated Nanobodies (Nbs) against S1 antigen were externally mixed with the tested samples (saliva or nasopharyngeal swab) before their application on the sample pad of the test strip containing angiotensin converting enzyme (ACE-2), as the capturing probe. The study included 320 individuals, with 180 being positively confirmed by RT-PCR and 140 confirmed negative, as well as, 45 health care workers, who were responsible for screening and handling of surgical cases in General Surgery Department and COVID clinic of TBRI. Our results proved that modified BS strip improved the overall sensitivity and specificity of S1antigen detection in saliva samples (95.21% and 99.29% respectively) compared to our previously developed enhanced LFTS (91.66% and 98.57% respectively). Also, the sensitivity of cases with Ct ≤ 30, Ct