Perovskite computed tomography imager and three-dimensional reconstruction

  • Yuhong He,
  • Jinmei Song,
  • Mingbian Li,
  • Kostiantyn Sakhatskyi,
  • Weijun Li,
  • Xiaopeng Feng,
  • Bai Yang,
  • Maksym Kovalenko,
  • Haotong Wei

摘要

Indirect scintillator computed tomography (CT) imaging suffers from reduced image contrast and high-dose X-ray exposure due to inevitable light losses and multiple energy conversion steps. Here we report a direct lead-halide perovskite CT imager through low-cost spray-coating processes. Detector arrays with 980 μm absorber thickness and <10 nm surface roughness yield uniform X-ray response with detection quantum efficiency of 80% and noise-equivalent dose of 153 pGyair. The perovskite CT imager affords the three-dimensional reconstruction of a tooth under a low effective dose of 5.5 μSv, about two orders of magnitude smaller than dental cone-beam CT, and low-contrast detectability by resolving a 5 Hounsfield unit difference within a 5 mm region of interest.