<p>Stroke volume, the volume of blood ejected by the left ventricle during a contraction, is a key metric of cardiovascular health. Currently, stroke volume is measured in clinic with specialised equipment. While purpose-made wearables exist to measure stroke volume, no solution relies solely on commodity devices. We present a deep learning system for stroke volume estimation from in-ear audio of earbuds. We combine generative self-supervised/transfer learning, a transformer-based autoencoder, to predict average stroke volume in unseen subjects. With data from 23 healthy participants, we compare our estimations to clinically validated device estimations. We achieve a mean absolute error of 5.24 ml, a Pearson correlation of r=0.94 between average predicted stroke volume and average true stroke volume, and a Percentage Error in the limits of agreement of 11.05% (within clinical range for stroke volume measurement devices). These findings open the doors to longitudinal, scalable and affordable cardiovascular measurement out of clinic.</p>

错误:搜索内容不能为空,请输入英文关键词
错误:关键词超出字数限制,请精简
高级检索

Measuring cardiac stroke volume through in-ear audio sensing

  • Kayla-Jade Butkow,
  • Navazh Jalaludeen,
  • Yang Liu,
  • Jake Stuchbury-Wass,
  • Qiang Yang,
  • Mathias Ciliberto,
  • Dong Ma,
  • Joseph Cheriyan,
  • Cecilia Mascolo

摘要

Stroke volume, the volume of blood ejected by the left ventricle during a contraction, is a key metric of cardiovascular health. Currently, stroke volume is measured in clinic with specialised equipment. While purpose-made wearables exist to measure stroke volume, no solution relies solely on commodity devices. We present a deep learning system for stroke volume estimation from in-ear audio of earbuds. We combine generative self-supervised/transfer learning, a transformer-based autoencoder, to predict average stroke volume in unseen subjects. With data from 23 healthy participants, we compare our estimations to clinically validated device estimations. We achieve a mean absolute error of 5.24 ml, a Pearson correlation of r=0.94 between average predicted stroke volume and average true stroke volume, and a Percentage Error in the limits of agreement of 11.05% (within clinical range for stroke volume measurement devices). These findings open the doors to longitudinal, scalable and affordable cardiovascular measurement out of clinic.