Background

Visual acuity (VA) represents a fundamental measure of visual function. The significant prevalence of underdiagnosed ocular disorders underscores the importance of effective VA assessment. This study evaluates the efficacy of a web-based VA assessment tool (“PocDoc”) versus conventional VA testing.

Methods

Prospective observational study including 353 participants recruited from various eye clinics in a tertiary referral centre. Age, diagnosis, and VA related information (i.e. VA measurements from PocDoc and conventional VA test [Snellen chart], test type, etc) were collected. Spearman’s rank correlation, Intraclass Correlation, and Bland-Altman plot compared outcomes of both tests. One-way ANOVA and paired-T test were used to compare means.

Results

Most patients were males (59.2%) with a mean age of 52.2 ± 20.6 years. PocDoc had moderate positive correlation to conventional testing (rho = 0.50, p < 0.001). PocDoc led to higher logMAR scores compared to conventional testing (mean logMAR 0.19 and 0.13 respectively, p < 0.01). Moreover, PocDoc demonstrated a sensitivity of 82.8% and specificity of 79% for detecting visual impairment. The discrepancy between PocDoc and conventional VA testing increased with higher logMAR values, indicating greater inconsistency between the tests for patients with poorer VA. Age, test type, and disease type contributed to this variability.

Conclusions

The concordance between PocDoc and conventional testing for VA measurement across various ages and conditions makes it a suitable screening tool. Future technological inventions should consider age, test type, and disease type as critical factors related to the level of agreement and correlation between digital and conventional VA testing methods.