错误:搜索内容不能为空,请输入英文关键词
错误:关键词超出字数限制,请精简
高级检索

Bilateral tibial fractures associated with powered exoskeleton use in complete spinal cord injury – a case report & literature review

  • John Mahon,
  • Lily Nolan,
  • David O’Sullivan,
  • Mark Curtin,
  • Aiden Devitt,
  • Colin G. Murphy

摘要

Introduction

Powered robotic exoskeleton (PRE) physiotherapy programmes are a relatively novel frontier which allow patients with reduced mobility to engage in supported walking. Research is ongoing regarding their utility, risks, and benefits. This article describes the case of two fractures occurring in one patient using a PRE.

Case

We report the case of a 54 year old man who sustained bilateral tibial fractures while using a PRE, on a background of T10 AIS A SCI. The initial session was discontinued due to acute severe bilateral knee swelling after approximately 15 min. The patient attended their local hospital the following day, where radiographs demonstrated bilateral proximal tibial fractures. The patient was treated with manipulation under anaesthetic and long-leg casting for five weeks, at which point he was stepped down to hinged knee braces which were weaned gradually while he remained non-weight bearing for 12 weeks. The patient was investigated with DEXA scan and was diagnosed with osteoporosis. He was liaised with rheumatology services and bone protection was initiated. Fracture healing was achieved and weight-bearing precautions were discontinued, however this period of immobilisation led to significant spasticity. The patient was discharged from orthopaedic services, with ongoing rehabilitation and physiotherapy follow-up.

Conclusion

PRE assisted physiotherapy programmes are a promising concept in terms of rehabilitation and independence, however they are not without risk and it is important that both providers and patients are aware of this. Furthermore, SCI patients are at increased risk for osteoporosis and should be monitored and considered for bone protection.