<p>Effective antiretroviral therapy has transformed HIV into a manageable chronic condition, provided viral suppression is maintained through routine viral load (VL) monitoring. Access to frequent VL testing remains limited, particularly outside centralized clinical settings. Decentralized and at-home HIV VL testing requires low-volume systems designed for patient operation. These systems must distinguish true viral suppression from test failure. Many approaches lack internal process verification, which makes negative results ambiguous. Here, we present ViraLite, an ultracompact, battery-powered HIV VL monitoring system that integrates reverse transcription loop-mediated isothermal amplification (RT-LAMP) with an RNase P internal process control, machine learning–assisted fluorescence analysis for one-pot multiplexing, and smartphone-guided operation. We evaluated ViraLite using 45 clinically archived plasma samples and benchmarked it against reverse transcription quantitative polymerase chain reaction (RT-qPCR). The internal process control identified 17 inconclusive tests that would otherwise be misclassified as negative. Among valid tests, ViraLite achieved 93.3% sensitivity and 100% specificity versus RT-qPCR. ViraLite enables decentralized HIV VL testing with low sample volume and interpretable negative results. This capability can expand monitoring beyond traditional clinic workflows and addresses a key barrier to patient-operated VL testing.</p><p></p>

错误:搜索内容不能为空,请输入英文关键词
错误:关键词超出字数限制,请精简
高级检索

ViraLite: an ultracompact HIV viral load self-testing system with internal quality control

  • Aneesh Kshirsagar,
  • Anthony J. Politza,
  • Tianyi Liu,
  • Md. Ahasan Ahamed,
  • Ming Dong,
  • Muhammad Asad Ullah Khalid,
  • Roland Jones,
  • Uttara Seshu,
  • Kathryn Risher,
  • Casey N. Pinto,
  • Yusheng Zhu,
  • Samir K. Gupta,
  • Weihua Guan

摘要

Effective antiretroviral therapy has transformed HIV into a manageable chronic condition, provided viral suppression is maintained through routine viral load (VL) monitoring. Access to frequent VL testing remains limited, particularly outside centralized clinical settings. Decentralized and at-home HIV VL testing requires low-volume systems designed for patient operation. These systems must distinguish true viral suppression from test failure. Many approaches lack internal process verification, which makes negative results ambiguous. Here, we present ViraLite, an ultracompact, battery-powered HIV VL monitoring system that integrates reverse transcription loop-mediated isothermal amplification (RT-LAMP) with an RNase P internal process control, machine learning–assisted fluorescence analysis for one-pot multiplexing, and smartphone-guided operation. We evaluated ViraLite using 45 clinically archived plasma samples and benchmarked it against reverse transcription quantitative polymerase chain reaction (RT-qPCR). The internal process control identified 17 inconclusive tests that would otherwise be misclassified as negative. Among valid tests, ViraLite achieved 93.3% sensitivity and 100% specificity versus RT-qPCR. ViraLite enables decentralized HIV VL testing with low sample volume and interpretable negative results. This capability can expand monitoring beyond traditional clinic workflows and addresses a key barrier to patient-operated VL testing.