错误:搜索内容不能为空,请输入英文关键词
错误:关键词超出字数限制,请精简
高级检索

High-frequency oscillatory ventilation with or without volume guarantee during neonatal transport

  • Vera Balog,
  • Gabor Liszkay,
  • Lajos Lantos,
  • Agnes Jermendy,
  • Gusztav Belteki

摘要

Objectives

To analyse deviation of ventilator parameters from their set targets during high-frequency oscillatory ventilation (HFOV) with or without volume guarantee (VG) and compare the two modes during emergency neonatal transport.

Study design

Retrospective observational study using the fabian™ HFOi ventilator.

Results

Median deviation of mean airway pressure from the set value was <1 cmH2O. During HFOV the pressure amplitude differed from the set value by <1 cmH2O. During HFOV-VG median deviation of the oscillation volume (VThf) from the targeted value was −0.07 mL/kg, but in some cases VThf was by >0.38 mL/kg below target. Setting maximum allowed amplitude 10 cmH2O above the usually required amplitude improved maintenance of VThf. HFOV and HFOV-VG parameters were similar, except the lower amplitude during HFOV without VG. VThf <2.5 mL/kg avoided hypercapnia in most cases.

Conclusions

HFOV and HFOV-VG maintain ventilator parameters close to their targets and are promising modalities during neonatal transport.