Comparison of dissolved iodine measurements in seawater between inductively coupled plasma mass spectrometry and voltammetry

  • Yuhi Satoh,
  • Yoshihito Ohtsuka

摘要

Two analytical methods, inductively coupled plasma mass spectrometry (ICP-MS) combined with high-performance liquid chromatography (HPLC) and voltammetry (VM), for three chemical species of dissolved iodine (iodide, iodate, and total dissolved iodine: TDI) were compared for dozens of coastal seawater samples owing to the compatibility of data between both methods. The median differences in the measured concentrations of TDI, total inorganic dissolved iodine (TII, the sum of iodide and iodate), and iodate between ICP-MS and VM were equivalent to 9.2, 13, and 14%, respectively. These differences were within the ranges that could be explained by the repeated-measurement precision of each measurement method for TDI, TII, and iodate. The difference for iodide was 19%, which was larger than the value based on the repeated-measurement precision for both methods. This is considered to be caused by the chemical instability and lower concentrations of iodide compared to other iodine species in seawater, in addition to the heterogeneity of natural samples. Finally, both methods provided reasonable measurement values for the iodine concentration in natural seawater samples.

Graphical abstract

Comparison of dissolved iodine measurements in seawater between inductively coupled plasma mass spectrometry and voltammetry