Adsorption enrichment integrated with paper-based devices for detection of trace levels of hexavalent chromium in water samples

  • Abdellah Muhammed,
  • Ahmed Hussen,
  • Takashi Kaneta

摘要

In the present study, a sensitive microfluidic paper-based analytical device (μ-PADs) integrated with adsorption enrichment procedure was developed to analyze Cr(VI) in water samples. The affecting factors, including pH and amounts of reagents were optimized. The limit of detection of 0.0015 mg L−1 and linear range of 0.005–2 mg L−1 were achieved with good intra- and inter-day precision of 5.1 and 7.6% RSD, respectively. The results obtained by the proposed method were validated by inductively coupled plasma-optical emission spectrometry (ICP-OES). The recoveries of the present method and ICP-OES were ranged from 96.3 to 109.0‬% and 106.0 to 109.7%, respectively. The two sets of (μ-PADs and ICP-OES) results were in a good agreement as paired t-test indicated no significant differences. The proposed method could be utilized for analyzing trace levels of Cr(VI) in water samples in the absence of conventional analytical instruments.