<p>The accurate detection of vital biomarkers such as Ascorbic Acid (AA), Uric Acid (UA) and Nitrite (NO<sub>2</sub><sup>−</sup>) is crucial for human health surveillance. However, existing methods often struggle with concurrent detection and quantification of multiple species, highlighting the need for a more effective solution. To address this challenge, this study aimed to develop a multifunctional electrochemical sensor capable of parallel detection of AA, UA and NO<sub>2</sub><sup>−</sup> using a synergistic combination of Graphene Oxide (GO) and Cadmium Sulfide (CdS) materials. Notably, the fabricated CdS@GO/Glassy Carbon Electrode (GCE) exhibited exceptional electrochemical activity, as evidenced by Differential Pulse Voltammetry (DPV) analysis. The sensor demonstrated remarkable sensitivity (8.13, 10.12, and 9.05 μA·μM<sup>−1</sup>·cm<sup>−2</sup>) and ultra-low detection limits (0.034, 0.062, and 0.084&#xa0;µM) for AA, UA and NO<sub>2</sub><sup>−</sup>, respectively. Furthermore, it successfully identified single molecules of each analyte in aqueous and biologic fluid samples, with recovery values comparable to those obtained using High-Performance Liquid Chromatography (HPLC) standard addition methods. The significance of this study lies in developing a novel CdS@GO/GCE sensor that enables concurrent detection and quantification of multiple vital biomarkers, offering a promising tool for human health monitoring and diagnosis.</p>

错误:搜索内容不能为空,请输入英文关键词
错误:关键词超出字数限制,请精简
高级检索

Facile fabrication of CdS@GO binary nanocomposite coated GCE for separate and parallel electrochemical sensing of ascorbic acid, uric acid and nitrite

  • A. Dhamodharan,
  • E. Murugan,
  • Huan Pang,
  • K. Perumal,
  • K. Jhansirani,
  • Yajun Gao

摘要

The accurate detection of vital biomarkers such as Ascorbic Acid (AA), Uric Acid (UA) and Nitrite (NO2) is crucial for human health surveillance. However, existing methods often struggle with concurrent detection and quantification of multiple species, highlighting the need for a more effective solution. To address this challenge, this study aimed to develop a multifunctional electrochemical sensor capable of parallel detection of AA, UA and NO2 using a synergistic combination of Graphene Oxide (GO) and Cadmium Sulfide (CdS) materials. Notably, the fabricated CdS@GO/Glassy Carbon Electrode (GCE) exhibited exceptional electrochemical activity, as evidenced by Differential Pulse Voltammetry (DPV) analysis. The sensor demonstrated remarkable sensitivity (8.13, 10.12, and 9.05 μA·μM−1·cm−2) and ultra-low detection limits (0.034, 0.062, and 0.084 µM) for AA, UA and NO2, respectively. Furthermore, it successfully identified single molecules of each analyte in aqueous and biologic fluid samples, with recovery values comparable to those obtained using High-Performance Liquid Chromatography (HPLC) standard addition methods. The significance of this study lies in developing a novel CdS@GO/GCE sensor that enables concurrent detection and quantification of multiple vital biomarkers, offering a promising tool for human health monitoring and diagnosis.