An Analysis of Input Parameters for Film-Based Flash X-Ray Radiography
摘要
Flash X-ray radiography (flash) is a commonly used diagnostic technique in dynamic experiments. An analysis of the effects of input parameters on resulting metrics of image quality can aid the experimentalist in configuring the X-ray input parameters to produce the highest quality radiograph for a given experiment. A flash X-ray test bed with HS800 film and a LANEX Medium F intensifier screen was used with an L3 450 kVp pulser and Scandiflash X-ray tube for this study. Input parameters including charge voltage, source filtering, and film-pack assembly were investigated for their impact on contrast-to-noise ratio (CNR), contrast, and contrast transfer function (CTF). Using VIDAR’s NDT Pro industrial film digitizer, scanner parameters such as optical density range, pixel spacing, scan mode, and digital bit-depth were also examined for their impact on image quality metrics. The highest CNR values were found with two LANEX intensifiers and no filtering. Charge voltage had no direct impact on CNR values. LANEX screen count and filtering resulted both in direct effects on CNR and interaction effects with each other and CNR value. Uncertainty bounds for CNR comparisons and repeatability of CTF evaluations are also discussed. Finally, the film results are compared with a previous study using other detector types, specifically Carestream INDUSTREX Flex GP, Flex HR, Flex XL Blue, and HPX-DR 3543.