Construction of a new multi-reflection time-of-flight mass spectrograph at RAON

  • Jun Young Moon,
  • Peter H. Schury,
  • Michiharu Wada,
  • Yoshikazu Hirayama,
  • Hiroari Miyatake,
  • Yutaka X. Watanabe,
  • Hironobu Ishiyama,
  • Marco Rosenbusch,
  • Yuta Ito,
  • Aiko Takamine,
  • Toshitaka Niwase,
  • Hermann Wollnik,
  • Zeren korkulu,
  • Taeksu Shin,
  • Jinho Lee

摘要

A new large-scale heavy-ion accelerator complex, RAON (Rare isotope Accelerator complex for ON-line experiments), in Korea has implemented the state-of-art device, multi-reflection time-of-flight mass spectrometer (MRTOF-MS) for the mass spectrometry of the exotic nuclei. Using the multi-reflections of ions between a pair of grid-free mirror electrodes and therefore extending the fly length, it can achieve the incredibly high mass resolving power of \(> 10^{5}\) > 10 5 within a short measurement time of < 10 ms, providing the nuclear mass of the exotic nuclei with a high-precision level of sub-ppm. The device is coupled to one of the beamlines in the ISOL experimental hall, ready for participating in the longstanding topics of nuclear structure in the outskirts of the nuclear chart and nucleosynthesis of heavy elements in explosive astrophysical environments. The exclusively high mass resolving power is also advantageous for reducing the isobaric contaminants that rare isotope ion beam facilities suffer from. In this paper, we present a summary of the construction and current performance of the RAON MRTOF-MS.