The verification of the utility of a commercially available phantom combination for quality control in contrast-enhanced mammography

  • J.-H. Kim,
  • M. Kessell,
  • D. Taylor,
  • M. Hill,
  • J. W. Burrage

摘要

Contrast-enhanced mammography is being increasingly implemented clinically, providing much improved contrast between tumour and background structures, particularly in dense breasts. Although CEM is similar to conventional mammography it differs via an additional exposure with high energy X-rays (≥ 40 kVp) and subsequent image subtraction. Because of its special operational aspects, the CEM aspect of a CEM unit needs to be uniquely characterised and evaluated. This study aims to verify the utility of a commercially available phantom set (BR3D model 020 and CESM model 022 phantoms (CIRS, Norfolk, Virginia, USA)) in performing key CEM performance tests (linearity of system response with iodine concentration and background subtraction) on two models of CEM units in a clinical setting. The tests were successfully performed, yielding results similar to previously published studies. Further, similarities and differences in the two systems from different vendors were highlighted, knowledge of which may potentially facilitate optimisation of the systems.