Potential dosimetric error in the adaptive workflow of a 1.5 T MR-Linac from patient movement relative to immobilisation systems
摘要
In magnetic resonance- (MR-) based adaptive workflows for an MR-linac, the treatment plan is optimized and recalculated online using the daily MR images. The Unity MR-linac is supplied with a patient positioning device (ppd) using pelvic and abdomen thermoplastic masks attached to a board with high-density components. This study highlights the dosimetric effect of using this in such workflows when there are relative patient-ppd displacements, as these are not visualized on MR imaging and the treatment planning system assumes the patient is fixed relative to the ppd. The online adapted plans of two example rectum cancer patients treated at a Unity MR-linac were perturbed by introducing relative patient-ppd displacements, and the effect was evaluated on plan dosimetry. Forty-eight perturbed clinical adapted plans were recalculated, based on online MR-based synthetic computed tomography, and compared with the original plans, using dose-volume histogram parameters and gamma analysis. The target volume covered by the prescribed dose (