Purpose <p>Cardiovascular phantoms are used in biomedical research and development applications, allowing for complex geometries to be studied in a controlled environment. The various layers of human tissue have been difficult to mimic in these phantoms. In this study, a novel dual-layer cardiovascular phantom is created.</p> Methods <p>The interior lumen is 3D printed using an elastic vat-photopolymerization resin and cast within an industry standard tissue-mimicking ballistics gel. Strips of the 3D-printed resin were prepared and tested to determine Young’s modulus, Ultimate tensile strength, and elongation at break.</p> Results <p>The final phantoms were reproducible, semi-transparent, and suitable for microCT scanning. Additionally, the 3D-printed elastic materials had: Young’s Modulus of 12 +/- 3.2&#xa0;MPa, UTS of 1.27 +/- 0.44&#xa0;MPa, and elongation at break of 29 +/- 9%. These results are within the physiological ranges of human tissues. There was a moderate correlation between the thickness of the sample and stiffness, which may be important depending on the application of the models.</p> Conclusion <p>The methods for producing a dual-layered phantom are reproducible and appropriate for a variety of biomedical applications.</p>

错误:搜索内容不能为空,请输入英文关键词
错误:关键词超出字数限制,请精简
高级检索

Development and Characteristics of a Dual-Layered Vascular Phantom

  • Kaitlyn M. Elmer,
  • Cassidy Caffin,
  • Breeanna Scott,
  • Sam E. Stephens,
  • Morten O. Jensen

摘要

Purpose

Cardiovascular phantoms are used in biomedical research and development applications, allowing for complex geometries to be studied in a controlled environment. The various layers of human tissue have been difficult to mimic in these phantoms. In this study, a novel dual-layer cardiovascular phantom is created.

Methods

The interior lumen is 3D printed using an elastic vat-photopolymerization resin and cast within an industry standard tissue-mimicking ballistics gel. Strips of the 3D-printed resin were prepared and tested to determine Young’s modulus, Ultimate tensile strength, and elongation at break.

Results

The final phantoms were reproducible, semi-transparent, and suitable for microCT scanning. Additionally, the 3D-printed elastic materials had: Young’s Modulus of 12 +/- 3.2 MPa, UTS of 1.27 +/- 0.44 MPa, and elongation at break of 29 +/- 9%. These results are within the physiological ranges of human tissues. There was a moderate correlation between the thickness of the sample and stiffness, which may be important depending on the application of the models.

Conclusion

The methods for producing a dual-layered phantom are reproducible and appropriate for a variety of biomedical applications.