Purpose <p>Atrial fibrillation (AF) is the most common chronic cardiac arrhythmia that increases the risk of stroke, primarily due to thrombus formation in the left atrial appendage (LAA). Left atrial appendage occlusion (LAAO) devices offer an alternative to oral anticoagulation for stroke prevention. However, the complex and variable anatomy of the LAA presents significant challenges to device design and deployment. Current benchtop models fail to replicate both anatomical variability and physiological hemodynamics, limiting their utility. This study introduces a novel left atrial cardiac simulator that incorporates patient-derived LAA models within a benchtop circulatory flow loop, enabling high-fidelity LAAO device testing and development.</p> Methods <p>A rigid, patient-derived left atrium (LA) model was 3D printed from segmented MRI data and modified to accommodate attachment of patient-specific LAA models. A library of LAA geometries was fabricated using silicone casting techniques to replicate the mechanical properties of native tissue. The LA-LAA model was integrated into a circulatory flow loop equipped with a pulsatile pump, pressure sensors, and flow probes, allowing real-time hemodynamic analysis. System tunability was demonstrated by varying heart rate, stroke volume, resistance, and compliance to simulate physiological and pathological conditions.</p> Results <p>The simulator accurately replicated LA pressure and flow waveforms, closely approximating physiological conditions. Changes in heart rate, stroke volume, and compliance effectively modulated LAP and LA inflow before and after LAAO. Distinct pressure and flow waveforms were observed with different LAA geometries. Hemodynamic analysis revealed increased left atrial pulse pressure after occlusion, with the greatest increase occurring after complete exclusion of the LAA. The simulator facilitated the evaluation of LAAO device performance, including metrics such as seal and PDL, and served as an effective training tool for iterative device deployment and recapture with visual and imaging-guided feedback.</p> Conclusions <p>The left atrial cardiac simulator offers a highly tunable and realistic platform for testing and developing LAAO devices. It also serves as an effective procedural training tool, allowing for the simulation of patient-specific anatomical and hemodynamic conditions. By enabling these advanced simulations, the simulator enhances pre-procedural planning, device sizing, and placement. This innovation represents a significant step toward advancing personalized medicine in atrial fibrillation management and improving LAAO outcomes.</p>

错误:搜索内容不能为空,请输入英文关键词
错误:关键词超出字数限制,请精简
高级检索

Design and Validation of a High-Fidelity Left Atrial Cardiac Simulator for the Study and Advancement of Left Atrial Appendage Occlusion

  • Keegan Mendez,
  • Manisha Singh,
  • Patrick Willoughby,
  • Beatrice Ncho,
  • Aileen Liao,
  • Susan Su,
  • Megan Lim,
  • Elijah Lee,
  • Mohamad Alkhouli,
  • Hasan Alarouri,
  • Ellen T. Roche

摘要

Purpose

Atrial fibrillation (AF) is the most common chronic cardiac arrhythmia that increases the risk of stroke, primarily due to thrombus formation in the left atrial appendage (LAA). Left atrial appendage occlusion (LAAO) devices offer an alternative to oral anticoagulation for stroke prevention. However, the complex and variable anatomy of the LAA presents significant challenges to device design and deployment. Current benchtop models fail to replicate both anatomical variability and physiological hemodynamics, limiting their utility. This study introduces a novel left atrial cardiac simulator that incorporates patient-derived LAA models within a benchtop circulatory flow loop, enabling high-fidelity LAAO device testing and development.

Methods

A rigid, patient-derived left atrium (LA) model was 3D printed from segmented MRI data and modified to accommodate attachment of patient-specific LAA models. A library of LAA geometries was fabricated using silicone casting techniques to replicate the mechanical properties of native tissue. The LA-LAA model was integrated into a circulatory flow loop equipped with a pulsatile pump, pressure sensors, and flow probes, allowing real-time hemodynamic analysis. System tunability was demonstrated by varying heart rate, stroke volume, resistance, and compliance to simulate physiological and pathological conditions.

Results

The simulator accurately replicated LA pressure and flow waveforms, closely approximating physiological conditions. Changes in heart rate, stroke volume, and compliance effectively modulated LAP and LA inflow before and after LAAO. Distinct pressure and flow waveforms were observed with different LAA geometries. Hemodynamic analysis revealed increased left atrial pulse pressure after occlusion, with the greatest increase occurring after complete exclusion of the LAA. The simulator facilitated the evaluation of LAAO device performance, including metrics such as seal and PDL, and served as an effective training tool for iterative device deployment and recapture with visual and imaging-guided feedback.

Conclusions

The left atrial cardiac simulator offers a highly tunable and realistic platform for testing and developing LAAO devices. It also serves as an effective procedural training tool, allowing for the simulation of patient-specific anatomical and hemodynamic conditions. By enabling these advanced simulations, the simulator enhances pre-procedural planning, device sizing, and placement. This innovation represents a significant step toward advancing personalized medicine in atrial fibrillation management and improving LAAO outcomes.