错误:搜索内容不能为空,请输入英文关键词
错误:关键词超出字数限制,请精简
高级检索

Microplate Versus Combined Microplate-Miniplate in Fixation of Zygomaticomaxillary Complex Fractures: An In-Silico Analysis of Biomechanical Parameters

  • Xaniar Mohammadi Khanaghah,
  • Tannaz Pourlak,
  • Ali Mortazavi

摘要

Introduction

To compare the biomechanical parameters of microplates and the combined miniplate-microplate for fixing zygomaticomaxillary complex (ZMC) fractures using nonlinear finite element analysis (FEA).

Material and Methods

Two samples of ZMC fracture models were prepared. In sample 1 (S1), the fractures were stabilized with microplates, and in sample 2 (S2), with miniplates plus microplates. FEA software was used to measure the displacement, Von Mises stress distribution (VMSD), and maximum principal stress distribution (MPSD).

Results

The displacement was 6.7 μm in the L-shaped plate of both samples, 4.4 μm in the S1 lateral-edge plate, 4.8 μm in the S2 lateral-edge plate, 5.8 μm in the S1 bottom-edge plate, and 5.6 μm in the S2 bottom-edge plate. The VMSD was 41.1 MPa in the S1 lateral-edge plate, 24.3 MPa in the S2 lateral-edge plate, 7.6 MPa in the S1 Lshaped plate, 9.6 MPa in S2 L-shaped plate, 28.5 MPa in the S1 bottom-edge plate, and 11.8 MPa in the S2 bottom-edge plate. The MPSD was 46.2 MPa in the S1 lateral-edge plate, 26.4 MPa in the S2 lateral-edge plate, 3.6 MPa in S1 L-shaped plate, 4.2 MPa in S2 L-shaped plate, 30.9 MPa in S1 bottom-edge plate, and 14.1 MPa in the S2 bottom-edge plate.

Conclusion

The L-shaped and lateral-edge plates in both samples had the highest and lowest amount of displacement, respectively. The lateral-edge plates in both samples had the highest VMSD and MPSD, which was higher in S1 than S2. The L-shaped plate had the lowest VMSD and MPSD in both samples.