Performance Evaluation of FPGA-based Standalone, Portable TDCR System

  • M. K. Sharma,
  • Shivam Agarwal,
  • M. S. Kulkarni,
  • Priyanka Reddy,
  • D. B. Kulkarni

摘要

The liquid scintillation counting based triple to double coincidence ratio (TDCR) method is a widely employed, direct activity determination technique to standardize pure beta emitters. An “FPGA-based standalone, portable TDCR system” is developed for absolute activity measurement of pure beta-emitting radionuclides. The system has a local 7″ touchscreen display to provide an intuitive GUI for operating the standalone instrument. The portable system will boost the capabilities of users in the field by providing a reference measurement method of radionuclide metrology. In nuclear medicine, the portable system will enable them to calibrate short-lived radionuclides within the dose-administering facility and avoid transporting radioactive sources to standardizing laboratories. The system’s performance is evaluated by comparing the activity determination results of 14C, 204Tl and 147Pm radionuclides with the ones from the CIEMAT/NIST method. The comparison shows that the activity measurement results from the system are closely aligned with those obtained using the CIEMAT/NIST method within the uncertainty limits. In this paper, the system’s performance evaluation is presented.