<p>Patient motion, particularly due to respiration, often introduces image distortions that compromise diagnostic accuracy in myocardial perfusion single-photon emission computed tomography (SPECT). To address this issue, we developed a novel respiratory motion reduction block (RRB) designed to minimize the respiratory motion of the heart. This study aims to evaluate the impact of the cardiac-centered with RRB (CC<sub>RRB</sub>) orbit, achieved using the RRB, on myocardial perfusion SPECT image quality. SPECT acquisition of a cardiac phantom was performed at the circular, neighboring elliptical (NE), and CC<sub>RRB</sub> orbits. The CC<sub>RRB</sub> orbit was achieved with RRB placed in front of the phantom based on the NE orbit. Count profile curves of the lesion and uniform slice images were obtained from the circumferential profile. Lesion contrast, normal accumulation uniformity, and count distortion were calculated from the circumferential profiles. Full width at half maximum (FWHM) was measured in the lateral, anterior, septal, and inferior walls of the myocardium, and both the mean and standard deviation (SD) were calculated. The lesion contrast was the highest in the NE orbit, slightly lower in the CC<sub>RRB</sub> orbit, and remarkably lower in the circular orbit than in the NE orbit. The uniformity and count distortion were superior for the CC<sub>RRB</sub> orbits. The SD of FWHM was greater in the circular and NE orbits. The CC<sub>RRB</sub> orbit effectively improves uniformity in SPECT imaging, preserving lesion contrast and spatial resolution. The CC<sub>RRB</sub> orbit provides a practical, accessible approach for enhancing image quality in clinical settings.</p>

错误:搜索内容不能为空,请输入英文关键词
错误:关键词超出字数限制,请精简
高级检索

Impact of a novel respiratory motion reduction block for myocardial perfusion SPECT image quality: a phantom study

  • Hajime Ichikawa,
  • Toyohiro Kato,
  • Takayuki Shibutani,
  • Mitsuaki Terabe,
  • Hideki Shimada

摘要

Patient motion, particularly due to respiration, often introduces image distortions that compromise diagnostic accuracy in myocardial perfusion single-photon emission computed tomography (SPECT). To address this issue, we developed a novel respiratory motion reduction block (RRB) designed to minimize the respiratory motion of the heart. This study aims to evaluate the impact of the cardiac-centered with RRB (CCRRB) orbit, achieved using the RRB, on myocardial perfusion SPECT image quality. SPECT acquisition of a cardiac phantom was performed at the circular, neighboring elliptical (NE), and CCRRB orbits. The CCRRB orbit was achieved with RRB placed in front of the phantom based on the NE orbit. Count profile curves of the lesion and uniform slice images were obtained from the circumferential profile. Lesion contrast, normal accumulation uniformity, and count distortion were calculated from the circumferential profiles. Full width at half maximum (FWHM) was measured in the lateral, anterior, septal, and inferior walls of the myocardium, and both the mean and standard deviation (SD) were calculated. The lesion contrast was the highest in the NE orbit, slightly lower in the CCRRB orbit, and remarkably lower in the circular orbit than in the NE orbit. The uniformity and count distortion were superior for the CCRRB orbits. The SD of FWHM was greater in the circular and NE orbits. The CCRRB orbit effectively improves uniformity in SPECT imaging, preserving lesion contrast and spatial resolution. The CCRRB orbit provides a practical, accessible approach for enhancing image quality in clinical settings.