Assessment of accuracy and repeatability of quantitative parameter mapping in MRI

  • Yuya Hirano,
  • Kinya Ishizaka,
  • Hiroyuki Sugimori,
  • Yo Taniguchi,
  • Tomoki Amemiya,
  • Yoshitaka Bito,
  • Kohsuke Kudo

摘要

We aimed to evaluate the accuracy and repeatability of the T1, T2*, and proton density (PD) values obtained by quantitative parameter mapping (QPM) using the ISMRM/NIST MRI system phantom and compared them with computer simulations. We compared the relaxation times and PD obtained through QPM with the reference values of the ISMRM/NIST MRI system phantom and conventional methods. Furthermore, we evaluated the presence or absence of influences other than noise in T1 and T2* values obtained by QPM by comparing the obtained coefficient of variation (CV) with simulation results. The T1, T2*, and PD values by QPM showed a strong correlation with the measured values and the referenced values. The simulated CVs of QPM calculated for each sphere showed similar trends to those of the actual scans.