Evaluation of calculation accuracy and computation time in a commercial treatment planning system for accelerator-based boron neutron capture therapy

  • Akihiko Takeuchi,
  • Katsumi Hirose,
  • Ryohei Kato,
  • Shinya Komori,
  • Mariko Sato,
  • Tomoaki Motoyanagi,
  • Yuhei Yamazaki,
  • Yuki Narita,
  • Yoshihiro Takai,
  • Takahiro Kato

摘要

This study aims to evaluate the feasibility of using a commercially available boron neutron capture therapy (BNCT) dose calculation program (NeuCure® Dose Engine) in terms of calculation accuracy and computation time. Treatment planning was simulated under the following calculation parameters: 1.5–5.0 mm grid sizes and 1–10% statistical uncertainties. The calculated monitor units (MUs) and computation times were evaluated. The MUs calculated on grid sizes larger than 2 mm were overestimated by 2% compared with the result of 1.5 mm grid. We established the two-step method for the routine administration of BNCT: multiple calculations involved in beam optimization should be done at a 5 mm grid and a 10% statistical uncertainty (the shortest computation time: 10.3 ± 2.1 min) in the first-step, and final dose calculations should be performed at a 2 mm grid and a 10% statistical uncertainty (satisfied clinical accuracy: 6.9 ± 0.3 h) in the second-step.