Purpose <p>Artificial intelligence (AI) algorithms for lung nodule detection assist radiologists. As their performance using ultra-high-resolution CT (U-HRCT) images has not been evaluated, we investigated the usefulness of 0.25-mm slices at U-HRCT using the commercially available deep-learning-based lung nodule detection (DL-LND) system.</p> Materials and methods <p>We enrolled 63 patients who underwent U-HRCT for lung cancer and suspected lung cancer. Two board-certified radiologists identified nodules more than 4&#xa0;mm in diameter on 1-mm HRCT slices and set the reference standard consensually. They recorded all lesions detected on 5-, 1-, and 0.25-mm slices by the DL-LND system. Unidentified nodules were included in the reference standard. To examine the performance of the DL-LND system, the sensitivity, and positive predictive value (PPV) and the number of false positive (FP) nodules were recorded.</p> Results <p>The mean number of lesions detected on 5-, 1-, and 0.25-mm slices was 5.1, 7.8 and 7.2 per CT scan. On 5-mm slices the sensitivity and PPV were 79.8% and 46.4%; on 1-mm slices they were 91.5% and 34.8%, and on 0.25-mm slices they were 86.7% and 36.1%. The sensitivity was significantly higher on 1- than 5-mm slices (<i>p</i> &lt; 0.01) while the PPV was significantly lower on 1- than 5-mm slices (<i>p</i> &lt; 0.01). A slice thickness of 0.25&#xa0;mm failed to improve its performance. The mean number of FP nodules on 5-, 1-, and 0.25-mm slices was 2.8, 5.2, and 4.7 per CT scan.</p> Conclusion <p>We found that 1&#xa0;mm was the best slice thickness for U-HRCT images using the commercially available DL-LND system.</p>

错误:搜索内容不能为空,请输入英文关键词
错误:关键词超出字数限制,请精简
高级检索

Performance of a deep-learning-based lung nodule detection system using 0.25-mm thick ultra-high-resolution CT images

  • Haruka Higashibori,
  • Wataru Fukumoto,
  • Sayaka Kusuda,
  • Kazushi Yokomachi,
  • Hidenori Mitani,
  • Yuko Nakamura,
  • Kazuo Awai

摘要

Purpose

Artificial intelligence (AI) algorithms for lung nodule detection assist radiologists. As their performance using ultra-high-resolution CT (U-HRCT) images has not been evaluated, we investigated the usefulness of 0.25-mm slices at U-HRCT using the commercially available deep-learning-based lung nodule detection (DL-LND) system.

Materials and methods

We enrolled 63 patients who underwent U-HRCT for lung cancer and suspected lung cancer. Two board-certified radiologists identified nodules more than 4 mm in diameter on 1-mm HRCT slices and set the reference standard consensually. They recorded all lesions detected on 5-, 1-, and 0.25-mm slices by the DL-LND system. Unidentified nodules were included in the reference standard. To examine the performance of the DL-LND system, the sensitivity, and positive predictive value (PPV) and the number of false positive (FP) nodules were recorded.

Results

The mean number of lesions detected on 5-, 1-, and 0.25-mm slices was 5.1, 7.8 and 7.2 per CT scan. On 5-mm slices the sensitivity and PPV were 79.8% and 46.4%; on 1-mm slices they were 91.5% and 34.8%, and on 0.25-mm slices they were 86.7% and 36.1%. The sensitivity was significantly higher on 1- than 5-mm slices (p < 0.01) while the PPV was significantly lower on 1- than 5-mm slices (p < 0.01). A slice thickness of 0.25 mm failed to improve its performance. The mean number of FP nodules on 5-, 1-, and 0.25-mm slices was 2.8, 5.2, and 4.7 per CT scan.

Conclusion

We found that 1 mm was the best slice thickness for U-HRCT images using the commercially available DL-LND system.