错误:搜索内容不能为空,请输入英文关键词
错误:关键词超出字数限制,请精简
高级检索

Efficacy and safety of inflator-monitored balloon-occluded retrograde transvenous obliteration for gastric varices

  • Dong Chan Lee,
  • Hyoung Nam Lee,
  • Youngjong Cho,
  • Sung-Joon Park,
  • Sangjoon Lee,
  • Seung Soo Kim,
  • Chan Ho Park,
  • Hyerim Park

摘要

Purpose

To investigate the diagnostic efficacy of inflator-monitored balloon-occluded retrograde transvenous obliteration (ImBRTO) in detecting balloon rupture and to evaluate the efficacy and safety of the ImBRTO in treating gastric varices.

Methods

Between June 2018 and June 2024, 31 consecutive patients (age: 61.7 ± 12.4 years; male 20, female 11) underwent ImBRTO for gastric varices. An occlusion balloon was inflated with an inflation device to monitor for balloon rupture during sclerosing, maintaining the balloon inflation for at least 3 h.

Results

The technical success rate was 100%. The incidence of balloon rupture was 6.5% (2/31), both diagnosed by a pressure drop in the inflation device. In one of the two patients, catheter replacement and an additional injection of a sclerosing agent were required due to early balloon rupture (< 3 h). Clinical success and complete elimination of gastric varices were achieved in all cases. No procedure-related complications were observed. There were no cases of recurrent variceal bleeding at the end of the follow-up (median: 5.2 months).

Conclusion

ImBRTO proves to be an effective and safe technique for treating gastric varices. The real-time monitoring of balloon integrity allows for timely decisions, resulting in excellent clinical outcomes.