Purpose: <p>Laparoscopic ultrasound (LUS) enhances the safety of liver surgery by visualizing intrahepatic vessels in real time. Still, vessel identification remains difficult due to probe constraints, complex vascular structure, and tissue deformation. This work aims to enable real-time, patient-specific vessel identification that remains robust under deformation through deformable ultrasound augmentation.</p> Methods: <p>Preoperative CT vessel annotations are used to generate synthetic ultrasound data via optimized physics-based rendering, coupled with domain adaptation to intraoperative ultrasound. The rendering is trained end to end for vessel identification and patient specificity, eliminating the need for preoperative ultrasound. A deformation-aware augmentation simulates realistic intraoperative motion and tissue deformation within the rendering pipeline.</p> Results: <p>In abdominal phantom and limited clinical feasibility experiments (single-case clinical evaluation), the framework achieved real-time intrahepatic vessel-branch identification, maintaining performance under new patient poses.</p> Conclusions: <p>The framework enables real-time vessel identification without preoperative ultrasound and supports technical feasibility, but multi-patient validation is still needed for generalizability and clinical feasibility.</p>

错误:搜索内容不能为空,请输入英文关键词
错误:关键词超出字数限制,请精简
高级检索

DefSynUS: Real-time patient-specific intrahepatic vessel identification via deformation-aware CT-US domain adaptation

  • Karl-Philippe Beaudet,
  • Yordanka Velikova,
  • Sidaty El Hadramy,
  • Nassir Navab,
  • Philippe Cattin,
  • Juan Verde,
  • Stéphane Cotin

摘要

Purpose:

Laparoscopic ultrasound (LUS) enhances the safety of liver surgery by visualizing intrahepatic vessels in real time. Still, vessel identification remains difficult due to probe constraints, complex vascular structure, and tissue deformation. This work aims to enable real-time, patient-specific vessel identification that remains robust under deformation through deformable ultrasound augmentation.

Methods:

Preoperative CT vessel annotations are used to generate synthetic ultrasound data via optimized physics-based rendering, coupled with domain adaptation to intraoperative ultrasound. The rendering is trained end to end for vessel identification and patient specificity, eliminating the need for preoperative ultrasound. A deformation-aware augmentation simulates realistic intraoperative motion and tissue deformation within the rendering pipeline.

Results:

In abdominal phantom and limited clinical feasibility experiments (single-case clinical evaluation), the framework achieved real-time intrahepatic vessel-branch identification, maintaining performance under new patient poses.

Conclusions:

The framework enables real-time vessel identification without preoperative ultrasound and supports technical feasibility, but multi-patient validation is still needed for generalizability and clinical feasibility.