Optimization of percutaneous intervention robotic system for skin insertion force

  • Benfang Duan,
  • Biao Jia,
  • Cheng Wang,
  • Shijia Chen,
  • Jun Xu,
  • Gao-Jun Teng

摘要

Purpose

Percutaneous puncture is a common interventional procedure, and its effectiveness is influenced by the insertion force of the needle. To optimize outcomes, we focus on reducing the peak force of the needle in the skin, aiming to apply this method to other tissue layers.

Methods

We developed a clinical puncture system, setting and measuring various variables. We analyzed their effects, introduced admittance control, set thresholds, and adjusted parameters. Finally, we validated these methods to ensure their effectiveness.

Results

Our system meets application requirements. We assessed the impact of various variables on peak force and validated the effectiveness of the new method. Results show a reduction of about 50% in peak force compared to the maximum force condition and about 13% compared to the minimum force condition. Finally, we summarized the factors to consider when applying this method.

Conclusion

To achieve peak force suppression, initial puncture variables should be set based on the trends in variable impact. Additionally, the factors of the new method should be introduced using these initial settings. When selecting these factors, the characteristics of the new method must also be considered. This process will help to better optimize peak puncture force.