From microscope to head-mounted display: integrating hand tracking into microsurgical augmented reality

  • Trishia El Chemaly,
  • Caio Athayde Neves,
  • Fanrui Fu,
  • Brian Hargreaves,
  • Nikolas H. Blevins

摘要

Purpose

The operating microscope plays a central role in middle and inner ear procedures that involve working within tightly confined spaces under limited exposure. Augmented reality (AR) may improve surgical guidance by combining preoperative computed tomography (CT) imaging that can provide precise anatomical information, with intraoperative microscope video feed. With current technology, the operator must manually interact with the AR interface using a computer. The latter poses a disruption in the surgical flow and is suboptimal for maintaining the sterility of the operating environment. The purpose of this study was to implement and evaluate free-hand interaction concepts leveraging hand tracking and gesture recognition as an attempt to reduce the disruption during surgery and improve human-computer interaction.

Methods

An electromagnetically tracked surgical microscope was calibrated using a custom 3D printed calibration board. This allowed the augmentation of the microscope feed with segmented preoperative CT-derived virtual models. Ultraleap’s Leap Motion Controller 2 was coupled to the microscope and used to implement hand-tracking capabilities. End-user feedback was gathered from a surgeon during development. Finally, users were asked to complete tasks that involved interacting with the virtual models, aligning them to physical targets, and adjusting the AR visualization.

Results

Following observations and user feedback, we upgraded the functionalities of the hand interaction system. User feedback showed the users’ preference for the new interaction concepts that provided minimal disruption of the surgical workflow and more intuitive interaction with the virtual content.

Conclusion

We integrated hand interaction concepts, typically used with head-mounted displays (HMDs), into a surgical stereo microscope system intended for AR in otologic microsurgery. The concepts presented in this study demonstrated a more favorable approach to human-computer interaction in a surgical context. They hold potential for a more efficient execution of surgical tasks under microscopic AR guidance.