Simulation of thoracic endovascular aortic repair in a perfused patient-specific model of type B aortic dissection

  • Lukas Mohl,
  • Roger Karl,
  • Matthias N. Hagedorn,
  • Armin Runz,
  • Stephan Skornitzke,
  • Malte Toelle,
  • C. Soeren Bergt,
  • Johannes Hatzl,
  • Christian Uhl,
  • Dittmar Böckler,
  • Katrin Meisenbacher,
  • Sandy Engelhardt

摘要

Purpose

Complicated type B Aortic dissection is a severe aortic pathology that requires treatment through thoracic endovascular aortic repair (TEVAR). During TEVAR a stentgraft is deployed in the aortic lumen in order to restore blood flow. Due to the complicated pathology including an entry, a resulting dissection wall with potentially several re-entries, replicating this structure artificially has proven to be challenging thus far.

Methods

We developed a 3d printed, patient-specific and perfused aortic dissection phantom with a flexible dissection flap and all major branching vessels. The model was segmented from CTA images and fabricated out of a flexible material to mimic aortic wall tissue. It was placed in a pulsatile hemodynamic flow loop. Hemodynamics were investigated through pressure and flow measurements and doppler ultrasound imaging. Surgeons performed a TEVAR intervention including stentgraft deployment under fluoroscopic guidance.

Results

The flexible aortic dissection phantom was successfully incorporated in the hemodynamic flow loop, a systolic pressure of 112 mmHg and physiological flow of 4.05 L per minute was reached. Flow velocities were higher in true lumen with a up to 35.7 cm/s compared to the false lumen with a maximum of 13.3 cm/s, chaotic flow patterns were observed on main entry and reentry sights. A TEVAR procedure was successfully performed under fluoroscopy. The position of the stentgraft was confirmed using CTA imaging.

Conclusions

This perfused in-vitro phantom allows for detailed investigation of the complex inner hemodynamics of aortic dissections on a patient-specific level and enables the simulation of TEVAR procedures in a real endovascular operating environment. Therefore, it could provide a dynamic platform for future surgical training and research.