OCT-based intra-cochlear imaging and 3D reconstruction: ex vivo validation of a robotic platform

  • Maarten Schoovaerts,
  • Mouloud Ourak,
  • Gianni Borghesan,
  • Tristan Putzeys,
  • Emmanuel Vander Poorten,
  • Nicolas Verhaert

摘要

Purpose

The small size of the cochlea, and its location deeply embedded in thick temporal bone, poses a challenge for intra-cochlear guidance and diagnostics. Current radiological imaging techniques are not able to visualize the cochlear microstructures in detail. Rotational optical coherence tomography (OCT) fibers show great potential for intra-cochlear guidance. The generated images could be used to map, and study, the tiny cochlear microstructures relevant for hearing.

Methods

This work describes the design of a rotational OCT probe with an outer diameter of 0.9 mm. It further discusses a robotic system, which features a remote center of motion mechanism, dedicated to the probe’s positioning, fine manipulation and stable insertion into the cochlear micro-spaces. Furthermore, the necessary calibration steps for 3D reconstruction are described, followed by a detailed quantitative analysis, comparing the 3D reconstructions using a synthetic, 2:1 scaled scala tympani model with a reconstruction from micro-CT, serving as the ground truth. Finally, the potential of the system is demonstrated by scanning a single ex vivo cadaveric human cochlea.

Results

The study investigates five insertions in the same 2:1 scaled tympani model, along with their corresponding 3D reconstruction. The comparison with micro-CT results in an average root-mean-square error of 74.2 µm, a signed distance error of 38.1 µm and a standard deviation of 63.6 µm. The average F-score of the reconstructions, using a distance threshold of 100 and 74.2 µm, resulted in 83.0% and 71.8%, respectively. Insertion in the cadaveric human cochlea showed the challenges for straight insertion, i.e., navigating the hook region.

Conclusion

Overall, the system shows great potential for intra-cochlear guidance and diagnostics, due to the system’s capability for precise and stable insertion into the basal turn in the scala tympani. The system, combined with the calibration procedure, results in detailed and precise 3D reconstructions.