Objective <p>To determine the diagnostic accuracy of cone-beam computed tomography (CBCT) in detecting simulated bony changes in the mandibular condyle by assessing the sensitivity and specificity.</p> Methods <p>This review adhered to PRISMA guidelines. Following predefined eligibility criteria, a search was conducted in four electronic databases in June 2024. The study-level risk of bias was assessed using a diagnostic test accuracy checklist provided by the Joanna Briggs Institute. Pooled estimates of sensitivity and specificity were calculated using a bivariate random-effects model.</p> Results <p>Among 1,803 potentially eligible references, six met the inclusion criteria for qualitative synthesis, and three&#xa0;for meta-analysis. The meta-analysis revealed that the index test, CBCT, had a low pooled sensitivity of 0.54&#xa0;and a high specificity of 0.93 for detecting simulated defects of the mandibular condyle. Computed tomography exhibited a lower sensitivity of 0.37,&#xa0;but&#xa0;similar specificity of 0.93&#xa0;like CBCT. Out of the six studies, five were found to have a low risk of bias.</p> Conclusions <p>Cone-beam computed tomography is found to be more accurate than other modalities for detecting condylar bony changes, effectively ruling out false positives, but with a risk of missing true positives. A smaller field of view and voxel size may provide more accurate detection.</p>

错误:搜索内容不能为空,请输入英文关键词
错误:关键词超出字数限制,请精简
高级检索

The diagnostic accuracy of cone-beam computed tomography for assessing in vitro osseous alterations of the mandibular condyle: a systematic review and meta-analysis

  • Nur Al Amin,
  • Taseef Hasan Farook,
  • Fahmi Oscandar,
  • Mohd Faizal Abdullah,
  • Norliza Binti Ibrahim,
  • Sumaiya Zabin Eusufzai,
  • Johari Yap Abdullah

摘要

Objective

To determine the diagnostic accuracy of cone-beam computed tomography (CBCT) in detecting simulated bony changes in the mandibular condyle by assessing the sensitivity and specificity.

Methods

This review adhered to PRISMA guidelines. Following predefined eligibility criteria, a search was conducted in four electronic databases in June 2024. The study-level risk of bias was assessed using a diagnostic test accuracy checklist provided by the Joanna Briggs Institute. Pooled estimates of sensitivity and specificity were calculated using a bivariate random-effects model.

Results

Among 1,803 potentially eligible references, six met the inclusion criteria for qualitative synthesis, and three for meta-analysis. The meta-analysis revealed that the index test, CBCT, had a low pooled sensitivity of 0.54 and a high specificity of 0.93 for detecting simulated defects of the mandibular condyle. Computed tomography exhibited a lower sensitivity of 0.37, but similar specificity of 0.93 like CBCT. Out of the six studies, five were found to have a low risk of bias.

Conclusions

Cone-beam computed tomography is found to be more accurate than other modalities for detecting condylar bony changes, effectively ruling out false positives, but with a risk of missing true positives. A smaller field of view and voxel size may provide more accurate detection.