错误:搜索内容不能为空,请输入英文关键词
错误:关键词超出字数限制,请精简
高级检索

High Sensitivity Meningococcus Love Wave Acoustic Biosensor

  • Chen Chen,
  • Yuting Wang,
  • Tinglun Ao,
  • Guopeng Hui,
  • Haotian Liu,
  • Keyu An

摘要

Early detection of meningococcus is critical to controlling its spread and saving lives in time. In this paper, a surface acoustic wave (SAW) device based on Electrode Width Control (EWC) IDT/PMMA/54°Y-X LiNbO3 is proposed to the detection of meningococcus by Love wave. The advantage of EWC IDT lies not only in overcoming the shortcomings of bidirectional and focused IDTs, but in improving the utilization of acoustic waves and the sensitivity of the sensor. The optimal waveguide layer thickness is determined to be λ/12 by comparing the S21 parameters with different waveguide layers thicknesses. On this basis, the Love wave sensor shows a sensitivity of 330 Hz (ng µL− 1)−1 and a detection limit of 50 pg µL− 1.