<p>Real-time, trace alpha activity monitoring in extreme mixed radiation fields is a well-known challenge in fission product-rich streams containing mixtures of beta-gamma-alpha radiation-emitting isotopes, arising due to the extreme 10<sup>6</sup>:1 ratio of background beta-gamma vs. alpha radiation activity levels. This paper presents results from a novel approach to addressing this challenge, based on Tensioned Metastable Fluid Detector (TMFD) sensing technology. An acoustically driven technique was developed and tested at Purdue University (using dissolved Rn-Po and U-chain alpha emitters), and validated (with dissolved Pu) at Savannah River National Laboratory for real-time monitoring of alpha activity in the 100–10,000 dpm/mL range.</p>

错误:搜索内容不能为空,请输入英文关键词
错误:关键词超出字数限制,请精简
高级检索

Real-time gross actinide monitoring in extreme (106:1) beta-gamma background fluid streams using mini-ATMFDs

  • Stepan Ozerov,
  • Nathan Boyle,
  • David DiPrete,
  • Rusi P. Taleyarkhan

摘要

Real-time, trace alpha activity monitoring in extreme mixed radiation fields is a well-known challenge in fission product-rich streams containing mixtures of beta-gamma-alpha radiation-emitting isotopes, arising due to the extreme 106:1 ratio of background beta-gamma vs. alpha radiation activity levels. This paper presents results from a novel approach to addressing this challenge, based on Tensioned Metastable Fluid Detector (TMFD) sensing technology. An acoustically driven technique was developed and tested at Purdue University (using dissolved Rn-Po and U-chain alpha emitters), and validated (with dissolved Pu) at Savannah River National Laboratory for real-time monitoring of alpha activity in the 100–10,000 dpm/mL range.