<p>This study demonstrates the potential of gamma spectrometry coupled with electrodeposition as a sample preparation technique for determination of <sup>55</sup>Fe in complex solid samples with high iron content. Electrodeposition has proven to be a reliable method for obtaining thin, homogeneous films containing up to 50&#xa0;mg of total iron. In contrast, ammonium pyrrolidinedithiocarbamate precipitation method shows inhomogeneity and significant film thickness at 0.5&#xa0;mg of iron. The efficiency correction function for gamma-spectrometric measurements based on the total iron content was introduced.</p>

错误:搜索内容不能为空,请输入英文关键词
错误:关键词超出字数限制,请精简
高级检索

Gamma spectrometry coupled with electrodeposition for 55Fe determination in solid samples containing significant amount of total iron

  • Katja Magdić Košiček,
  • Ljudmila Benedik,
  • Marijan Nečemer,
  • Branko Vodenik,
  • Benjamin Zorko,
  • Željko Grahek,
  • Ivana Tucaković

摘要

This study demonstrates the potential of gamma spectrometry coupled with electrodeposition as a sample preparation technique for determination of 55Fe in complex solid samples with high iron content. Electrodeposition has proven to be a reliable method for obtaining thin, homogeneous films containing up to 50 mg of total iron. In contrast, ammonium pyrrolidinedithiocarbamate precipitation method shows inhomogeneity and significant film thickness at 0.5 mg of iron. The efficiency correction function for gamma-spectrometric measurements based on the total iron content was introduced.