<p>The hardware of the pneumatic transfer system used to perform instrumental neutron activation analysis at the IBR-2 reactor (JINR, Dubna) has undergone a modernization. For this purpose, the new control panel, representing a desktop program for Windows, was developed. In terms of the gamma spectrometry automation outdated sample changers were replaced by four KUKA KR10 R1100 manipulators with all necessary controllers. Additionally, to use fixed geometry of irradiated samples during induced activity measurement special containers and a protective box for sample repacking were developed.</p>

错误:搜索内容不能为空,请输入英文关键词
错误:关键词超出字数限制,请精简
高级检索

Modernization on the Regata facility (IBR-2 reactor) designed for instrumental neutron activation analysis

  • Dmitrii Grozdov,
  • Vladimir Galustov,
  • Inga Zinicovscaia

摘要

The hardware of the pneumatic transfer system used to perform instrumental neutron activation analysis at the IBR-2 reactor (JINR, Dubna) has undergone a modernization. For this purpose, the new control panel, representing a desktop program for Windows, was developed. In terms of the gamma spectrometry automation outdated sample changers were replaced by four KUKA KR10 R1100 manipulators with all necessary controllers. Additionally, to use fixed geometry of irradiated samples during induced activity measurement special containers and a protective box for sample repacking were developed.