<p>In this paper, a method and its device were designed to quickly adjust the effective decay constant <i>λ</i><sub>EFF</sub> of a radon exhalation simulation device. The device achieves the purpose of adjusting<i> λ</i><sub>EFF</sub> by actively adjusting the leakage rate of the device. The <i>λ</i><sub>EFF</sub> of the initial radon exhalation simulation device was measured firstly. then five different sets of <i>λ</i><sub>EFF</sub> adjustment experiments were set. The error between the actual value of <i>λ</i><sub>EFF</sub> and the target value of<i> λ</i><sub>EFF</sub> calculated from the five experiments was within 2 standard deviations. This shows that the method of adjusting <i>λ</i><sub><i>E</i>FF</sub> of the device is correct.</p>

错误:搜索内容不能为空,请输入英文关键词
错误:关键词超出字数限制,请精简
高级检索

A method for adjusting the effective decay constant of a radon exhalation simulation device

  • Zhipeng Liu,
  • Tao Hu,
  • Zhongkai Fan,
  • Huiying Li,
  • Jiale Sun,
  • Xinyue Yang,
  • Jia Liu,
  • Haibo Yi,
  • Jiulin Wu,
  • Hongzhi Yuan,
  • Yanliang Tan

摘要

In this paper, a method and its device were designed to quickly adjust the effective decay constant λEFF of a radon exhalation simulation device. The device achieves the purpose of adjusting λEFF by actively adjusting the leakage rate of the device. The λEFF of the initial radon exhalation simulation device was measured firstly. then five different sets of λEFF adjustment experiments were set. The error between the actual value of λEFF and the target value of λEFF calculated from the five experiments was within 2 standard deviations. This shows that the method of adjusting λEFF of the device is correct.