错误:搜索内容不能为空,请输入英文关键词
错误:关键词超出字数限制,请精简
高级检索

Direct determination of the 237Np/238U ratio in nuclear fuel samples by multicollection inductively coupled plasma mass spectrometry (MC-ICP-MS)

  • S. Mialle,
  • A. Nonell,
  • C. Cruchet,
  • C. Caussignac,
  • M. Marie,
  • A. Rincel,
  • H. Isnard

摘要

A new method for the direct measurement of 237Np/238U ratio in irradiated UO2 pellets by multicollection inductively coupled plasma mass spectrometry (MC-ICP-MS) is proposed. It allows the determination of ratios down to 10 × 10–6 mol·mol−1 using ion counter and Faraday cup. This approach was validated by intercomparison with the usual two-step-method (Quadrupole ICP-MS for 237Np determination and isotope dilution mass spectrometry (IDMS) for 238U). For ratios between 10 × 10–6 and 100 × 10–6 mol·mol−1, expanded uncertainties (k = 2) varied from 2.75% to 0.81%, twice lower than the uncertainties determined by the usual method.