Electro-Thermal Simulation and Evaluation of Transition Edge Sensor X-ray Microcalorimeter with Mushroom-Type Absorber

  • Ryo Ota,
  • Keita Tanaka,
  • Tasuku Hayashi,
  • Rikuta Miyagawa,
  • Yuta Yagi,
  • Noriko Y. Yamasaki,
  • Kazushisa Mitsuda

摘要

High-efficiency microcalorimeter arrays are required as imaging spectroscopy for X-ray microscopy in general and for astrophysics. We have developed in-house an array of Transition Edge Sensor (TES) X-ray microcalorimeters with mushroom-type absorbers which have a large area of \(260\,\upmu \hbox {m}\) 260 μ m square with multiple support stems. Insufficient thermal conductivity may degrade the energy resolution. We construct a 3D FEM simulation with thermal conduction and electric feedback circuit of TES. Estimated physical parameters were incorporated into the simulation to reproduce the measured pulse waveforms. As a result, the effect of the large absorber on the degradation of energy resolution was small for the mushroom-type absorber TES calorimeter.