<p>We developed a smartphone application (SMART-QT application) that can semi-automatically measure QT and QTc intervals based on a snapshot of the electrocardiogram (ECG) trace and the heart rate displayed on a patient monitor. In this study, we aimed to validate the SMART-QT application. In this prospective single-center method comparison study, we measured QT and QTc intervals with the SMART-QT application (QT<sub>APP</sub> and QTc<sub>APP</sub>; test method) and simultaneously manually measured QT and QTc intervals from a 12-lead ECG (QT<sub>REF</sub> and QTc<sub>REF</sub>; reference method) in 57 adult volunteers and patients who had sinus rhythm and no acute or chronic cardiac comorbidities. To investigate the agreement between QT<sub>APP</sub> and QT<sub>REF</sub> and between QTc<sub>APP</sub> and QTc<sub>REF</sub>, we performed Bland-Altman analyses and calculated the mean of the differences, the standard deviation, and the 95%-limits of agreement (95%-LOA). We defined clinically acceptable agreement as maximum mean of the differences ± standard deviation of 20 ± 20 ms. The mean of the differences between QT<sub>APP</sub> and QT<sub>REF</sub> was 14 ± 20 ms (95%-LOA −26 to 54 ms). The mean of the differences between QTc<sub>APP</sub> and QTc<sub>REF</sub> was 13 ± 15 ms (95%-LOA −16 to 42 ms). The agreement between QT<sub>APP</sub> and QT<sub>REF</sub> and between QTc<sub>APP</sub> and QTc<sub>REF</sub> was clinically acceptable in adult volunteers and patients without cardiac comorbidities.</p>

错误:搜索内容不能为空,请输入英文关键词
错误:关键词超出字数限制,请精简
高级检索

A smartphone application for semi-automated QT interval analysis based on a snapshot of an electrocardiogram trace displayed on a patient monitor

  • David Beckmann,
  • Moritz Flick,
  • Karim Kouz,
  • Bernd Saugel

摘要

We developed a smartphone application (SMART-QT application) that can semi-automatically measure QT and QTc intervals based on a snapshot of the electrocardiogram (ECG) trace and the heart rate displayed on a patient monitor. In this study, we aimed to validate the SMART-QT application. In this prospective single-center method comparison study, we measured QT and QTc intervals with the SMART-QT application (QTAPP and QTcAPP; test method) and simultaneously manually measured QT and QTc intervals from a 12-lead ECG (QTREF and QTcREF; reference method) in 57 adult volunteers and patients who had sinus rhythm and no acute or chronic cardiac comorbidities. To investigate the agreement between QTAPP and QTREF and between QTcAPP and QTcREF, we performed Bland-Altman analyses and calculated the mean of the differences, the standard deviation, and the 95%-limits of agreement (95%-LOA). We defined clinically acceptable agreement as maximum mean of the differences ± standard deviation of 20 ± 20 ms. The mean of the differences between QTAPP and QTREF was 14 ± 20 ms (95%-LOA −26 to 54 ms). The mean of the differences between QTcAPP and QTcREF was 13 ± 15 ms (95%-LOA −16 to 42 ms). The agreement between QTAPP and QTREF and between QTcAPP and QTcREF was clinically acceptable in adult volunteers and patients without cardiac comorbidities.