Background <p>Catheter ablation using radiofrequency (RF) energy is an established treatment for ventricular tachycardia (VT). Tissue temperature is a key determinant of successful lesion creation, and yet, it is difficult to measure during conventional RF ablation because of the cooling effect of high-flow rate saline irrigation. The TRAC-VT study evaluated the safety and efficacy of a novel irrigated RF ablation system modulating power based on real-time tissue temperature.</p> Methods <p>Patients with sustained monomorphic VT and structural heart disease (SHD) were enrolled. Catheter ablation was performed in temperature-control mode (irrigation 8 ml/min, temperature set-points 55 or 60 °C, and power output ≤ 50 W), with RF applications for ≤ 45 s. The primary safety endpoint was a composite of cardiovascular-specific serious procedure-related adverse events within 30 days post-ablation. The primary effectiveness endpoint was acute success (i.e., non-inducibility of all clinically relevant VTs).</p> Results <p>Thirty-eight patients were enrolled with monomorphic VT (age 68 ± 12 years and 84% male), with an average of 1.7 ± 1.2 VTs targeted per patient. In total, 41 ± 23 RF applications per patient were delivered. Acute procedural success was 100% (95% CI, 91–100%). No primary safety endpoints were observed. Six-month follow-up was completed in 92% of patients with 81% (95% CI, 65–91%) freedom from sustained or treated VT. A repeat ablation was performed in three patients.</p> Conclusions <p>Ablation of VT in SHD, using a temperature-controlled irrigated RF catheter, was safe and effective with a low rate of VT recurrence at 6 months.</p> Graphical abstract <p></p>

错误:搜索内容不能为空,请输入英文关键词
错误:关键词超出字数限制,请精简
高级检索

Safety and efficacy of a temperature-controlled ablation system for ventricular tachycardia: Results from the TRAC-VT study

  • Josef Kautzner,
  • Javier Moreno,
  • Claudio Tondo,
  • Frédéric Anselme,
  • James Burrell,
  • Daniel Becker,
  • Petr Peichl,
  • Ian Patchett,
  • Tarvinder Dhanjal

摘要

Background

Catheter ablation using radiofrequency (RF) energy is an established treatment for ventricular tachycardia (VT). Tissue temperature is a key determinant of successful lesion creation, and yet, it is difficult to measure during conventional RF ablation because of the cooling effect of high-flow rate saline irrigation. The TRAC-VT study evaluated the safety and efficacy of a novel irrigated RF ablation system modulating power based on real-time tissue temperature.

Methods

Patients with sustained monomorphic VT and structural heart disease (SHD) were enrolled. Catheter ablation was performed in temperature-control mode (irrigation 8 ml/min, temperature set-points 55 or 60 °C, and power output ≤ 50 W), with RF applications for ≤ 45 s. The primary safety endpoint was a composite of cardiovascular-specific serious procedure-related adverse events within 30 days post-ablation. The primary effectiveness endpoint was acute success (i.e., non-inducibility of all clinically relevant VTs).

Results

Thirty-eight patients were enrolled with monomorphic VT (age 68 ± 12 years and 84% male), with an average of 1.7 ± 1.2 VTs targeted per patient. In total, 41 ± 23 RF applications per patient were delivered. Acute procedural success was 100% (95% CI, 91–100%). No primary safety endpoints were observed. Six-month follow-up was completed in 92% of patients with 81% (95% CI, 65–91%) freedom from sustained or treated VT. A repeat ablation was performed in three patients.

Conclusions

Ablation of VT in SHD, using a temperature-controlled irrigated RF catheter, was safe and effective with a low rate of VT recurrence at 6 months.

Graphical abstract