Pulsed-field ablation for atrial fibrillation without the use of fluoroscopy

  • Nicholas O. Palmeri,
  • Daniel Alyesh,
  • Madison Keith,
  • Erin Greenhaw,
  • Cooper Erickson,
  • William Choe,
  • Sri Sundaram

摘要

Introduction

Pulsed-field ablation (PFA) and fluoroless ablation (FA) are emerging techniques in contemporary in electrophysiology. With widespread use of 3D electroanatomic mapping systems and advanced intracardiac echo (ICE) imaging, fluoroless ablation has become more widely adopted. However, with the importance of tissue contact for lesion durability, initial PFA has been used with fluoroscopic guidance, but both ICE and electroanatomic mapping make fluoroless PFA feasible. The objective of this study is to demonstrate that PFA can be done safely and effectively without fluoroscopy.

Methods

At a single center, consecutive patients undergoing ablation with a pentaspline PFA catheter using a fluoroless approach are described. The standard 3D anatomic map settings were adjusted with changes in interior and exterior projection, respiratory compensation, and interpolation. In addition, projection map lesions were used to confirm adequate circumferential ablation lesions. ICE was used extensively for wire guidance and evaluation of contact with tissue.

Results

Beginning on March 15, 2024, 50 consecutive subjects (19 female/31 male) aged 68.0 (± 13.7) underwent PFA ablation. The average CHA2DS2-VA2Sc score was 3.0 (± 1.9). The average LVEF was 57.3% (± 10.0) and the average LA size was 3.9 cm (± 1.2). Projection lesions were placed with every application of PFA. An average of 41.7 (± 8.5) PFA applications were placed. In 100% (50/50) of subjects, acute isolation of the pulmonary veins was achieved. Eighteen subjects also underwent concomitant posterior wall isolation and in 100% of these subjects, posterior isolation was achieved. There were zero complications in this cohort. In 50/50 subjects (100%), fluoroscopy was not used. In comparison to the control cohort, the LA dwell time of the ablation catheter was similar (p = 0.34).

Conclusion

In comparison to the traditional PFA with fluoroscopy, this proof-of-concept study shows fluoroless PFA ablation can be performed safely and with similar acute success rates as with use of fluoroscopy.