Clinical outcomes of conduction system pacing compared to biventricular pacing in patients with mid-range ejection fraction
摘要
There is a paucity of data comparing conduction system pacing (CSP) to biventricular pacing (BiVP) in patients with heart failure (HF) with mid-range left ventricular ejection fraction (LVEF).
ObjectiveCompare the clinical outcomes of patients with mid-range LVEF undergoing CSP versus BiVP.
MethodsPatients with mid-range LVEF (> 35 to 50%) undergoing CSP or BiVP were retrospectively identified. Lead performance, LVEF, HF hospitalization, and clinical composite outcome including upgrade to cardiac resynchronization therapy and mortality were compared.
ResultsA total of 36 patients (20 BiVP, 16 CSP--14 His bundle pacing, 4 left bundle branch area pacing) were analyzed. The mean age was 73 ± 15, 44% were female, and the mean LVEF was 42 ± 5%. Procedural and fluoroscopy time was comparable between the two groups. QRS duration was significantly shorter for the CSP group compared to the BiVP group (P < 0.001). During a mean follow-up of 47 ± 36 months, no significant differences were found in thresholds or need for generator change due to early battery depletion. LVEF improved in both groups (41.5 ± 4.5% to 53.9 ± 10.9% BiVP, P < 0.001; 41.6 ± 5.3% to 52.5 ± 8.3% CSP, P < 0.001). There were no significant differences in HF hospitalizations (P = 0.71) or clinical composite outcomes (P = 0.07).
ConclusionAmong patients with HF with moderately reduced ejection fraction, CSP appears associated with similar improvements in LVEF and had similar clinical outcomes as BiVP in mid-term follow-up.