Purpose <p>Ultrasound measurement of follicle diameter is essential in IVF monitoring. This study evaluates the analytical performance of follicle counts and size measurements from two-dimensional images using an AI-based platform, compared to assessments by certified sonographers.</p> Methods <p>A total of 5508 TVUS scans from 1689 patients undergoing controlled ovarian stimulation across four IVF centers (Poland, Argentina, Colombia, and the USA) were retrospectively analyzed. All visible follicles were marked using bounding boxes. The dataset included three subsets: training/validation for model development, independent test for evaluating performance across ultrasound systems, and a consensus test set (102 scans from 27 patients) annotated by three expert sonographers. Model performance was assessed using precision, recall, and F1 score. Annotation efficiency was measured by comparing manual and AI-assisted times. Real-world performance was evaluated on a prospective cohort of 904 scans from 269 patients, based on expert adjustments to AI annotations.</p> Results <p>For follicles ≥ 10&#xa0;mm, the model achieved 98.2% precision (95% CI, 96.5–99.2), 88.9% recall (85.0–91.8), and 93.3% F1 score (90.7–95.1). For all follicles, precision and recall were 94.2% (92.8–95.4) and 68.9% (65.9–71.9). Annotation time was reduced 2.5-fold (<i>p</i> &lt; 0.01), with an average of 0.54 expert adjustments per scan (CI, 0.47–0.62). Model performance was stable across ultrasound platforms.</p> Conclusion <p>This AI platform enables accurate, automated follicle counting and measurement during ovarian stimulation. It matches expert-level performance, improves efficiency, and supports scalable, cost-effective fertility care without compromising quality.</p>

错误:搜索内容不能为空,请输入英文关键词
错误:关键词超出字数限制,请精简
高级检索

An artificial intelligence platform for automated measurement and count estimation of ovarian follicles during ovarian stimulation and IVF: a multicenter study

  • Piotr Wygocki,
  • Andrzej Zapała,
  • Mateusz Ulfig,
  • Marcin Zieleń,
  • Krystian Zieliński,
  • Natalia Gajewska,
  • Damian Drzyzga,
  • Marcin Wrochna,
  • Piotr Sankowski,
  • Gerard Letterie

摘要

Purpose

Ultrasound measurement of follicle diameter is essential in IVF monitoring. This study evaluates the analytical performance of follicle counts and size measurements from two-dimensional images using an AI-based platform, compared to assessments by certified sonographers.

Methods

A total of 5508 TVUS scans from 1689 patients undergoing controlled ovarian stimulation across four IVF centers (Poland, Argentina, Colombia, and the USA) were retrospectively analyzed. All visible follicles were marked using bounding boxes. The dataset included three subsets: training/validation for model development, independent test for evaluating performance across ultrasound systems, and a consensus test set (102 scans from 27 patients) annotated by three expert sonographers. Model performance was assessed using precision, recall, and F1 score. Annotation efficiency was measured by comparing manual and AI-assisted times. Real-world performance was evaluated on a prospective cohort of 904 scans from 269 patients, based on expert adjustments to AI annotations.

Results

For follicles ≥ 10 mm, the model achieved 98.2% precision (95% CI, 96.5–99.2), 88.9% recall (85.0–91.8), and 93.3% F1 score (90.7–95.1). For all follicles, precision and recall were 94.2% (92.8–95.4) and 68.9% (65.9–71.9). Annotation time was reduced 2.5-fold (p < 0.01), with an average of 0.54 expert adjustments per scan (CI, 0.47–0.62). Model performance was stable across ultrasound platforms.

Conclusion

This AI platform enables accurate, automated follicle counting and measurement during ovarian stimulation. It matches expert-level performance, improves efficiency, and supports scalable, cost-effective fertility care without compromising quality.