In Vitro and In Silico Characterization of the Aggregation of Thrombi on Textured Ventricular Cannula
摘要
The unacceptably high stroke rate associated with HeartMate 3 ventricular assist device (VAD) without signs of adherent pump thrombosis is hypothesized to be the result of the emboli produced by the inflow cannula, that are ingested and ejected from the pump. This in vitro and numerical study aimed to emulate the surface features and supraphysiological shear of a ventricular cannula to provide insight into their effect on thrombogenesis. Human whole blood was perfused at calibrated flow rates in a microfluidic channel to achieve shear rates 1000–7500 s−1, comparable to that experienced on the cannula. The channel contained periodic teeth representative of the rough sintered surface of the HeartMate 3 cannula. The deposition of fluorescently labeled platelets was visualized in real time and analyzed with a custom entity tracking algorithm. Numerical simulations of a multi-constituent thrombosis model were performed to simulate laminar blood flow in the channel. The sustained growth of adherent platelets was observed in all shear conditions (