Examination of rapid adjustment system based on screen score obtained using continuous shear wave elastography

  • Marie Tabaru,
  • Ren Koda,
  • Hitoshi Shitara,
  • Hirotaka Chikuda,
  • Yoshiki Yamakoshi

摘要

Purpose

Continuous shear wave elastography (C-SWE) can be expected to be applied to portable muscle elasticity diagnosis. To establish diagnostic technology, it will be necessary to improve measurement techniques and quantitative measurement accuracy.

Methods

In this study, we investigated two screen scores: the quality index (Q-index), which determines whether the intensity of a power Doppler image is appropriate, and the shear wave propagation direction index (SWDI), which determines the uniformity of shear wave propagation.

Results

First, we performed numerical simulations with white noise and found that the coefficient of variation of shear wave velocity estimation was less than 5% when the normalized Q-index was greater than 0.27. Furthermore, regarding the SWDI, we clarified the relationship between the standard deviation in shear wave propagation direction and the SWDI. Next, the relationship between the Q-index and coefficient of variation of estimated shear wave velocity was evaluated through experiments using a tissue-mimicking phantom. The results showed that there was a negative correlation between the Q-index and the coefficient of variation, and the fluctuation of the propagation velocity could be inferred from the Q-index. Finally, we showed the results of applying the screen scores to muscle relaxation monitoring and confirmed its usefulness in clinical applications.

Conclusion

By applying the screen scores, we showed improved stability in speed estimation in C-SWE, and demonstrated the possibility of clinical applicability.