Purpose <p>To compare the efficac and safety of a dual-blade 20,000 cuts per minute (cpm) vitrectomy probe with a single-blade 10,000&#xa0;cpm probe for primary rhegmatogenous retinal detachment (RRD).</p> Study design <p>Prospective, randomized controlled clinical trial.</p> Methods <p>Evaluations were conducted preoperatively, intraoperatively, and at three months postoperatively. The main outcome measure focused on the duration of core vitrectomy, with secondary outcomes including peripheral vitreous shaving duration, balanced salt solution (BSS) consumption, anatomical and functional outcomes, as well as the incidence of surgical procedure-related complications.</p> Results <p>Overall, 35 cases in the 20,000&#xa0;cpm group and 37 cases in the 10,000&#xa0;cpm group completed the trial without significant differences in baseline demographic characteristics. The 20,000&#xa0;cpm group demonstrated significantly shorter core vitrectomy duration (161.6 ± 10.4 vs. 206.8 ± 10.1&#xa0;s) (<i>P</i> = 0.003) and peripheral vitreous shaving time (446.3 ± 20.3 vs. 544.2 ± 22.2&#xa0;s) (<i>P</i> = 0.002) compared to the 10,000&#xa0;cpm group. BSS consumption was higher in the 20,000&#xa0;cpm group but without statistical significance (<i>P</i> = 0.231). There were no significant differences in the need for scleral sutures and the incidence of iatrogenic retinal breaks (<i>P</i> = 0.331 and 0.523). At the 3-month follow-up, there were no statistically significant differences in primary success, final anatomical success, or mean visual acuity (<i>P</i> &gt; 0.9, <i>P</i> = 0.326). Rates of complications, including ocular hypertension, epiretinal membrane formation, and endophthalmitis also showed no statistically significant differences (<i>P</i> &gt; 0.6).</p> Conclusion <p>The utilization of the new-generation 27-gauge system with a 20,000&#xa0;cpm probe may herald a new paradigm of high-flow, smaller-diameter instrumentation, thereby enhancing the efficiency of the small gauge technique.</p>

错误:搜索内容不能为空,请输入英文关键词
错误:关键词超出字数限制,请精简
高级检索

Efficiency and safety comparison between 20,000 and 10,000 cuts per minute vitrectomy using a 27G cutter in rhegmatogenous retinal detachment: a prospective randomized controlled study

  • Yu-Te Huang,
  • I. Wang,
  • Po-Yu Chen,
  • Chi-Lan Yang,
  • San-Ni Chen

摘要

Purpose

To compare the efficac and safety of a dual-blade 20,000 cuts per minute (cpm) vitrectomy probe with a single-blade 10,000 cpm probe for primary rhegmatogenous retinal detachment (RRD).

Study design

Prospective, randomized controlled clinical trial.

Methods

Evaluations were conducted preoperatively, intraoperatively, and at three months postoperatively. The main outcome measure focused on the duration of core vitrectomy, with secondary outcomes including peripheral vitreous shaving duration, balanced salt solution (BSS) consumption, anatomical and functional outcomes, as well as the incidence of surgical procedure-related complications.

Results

Overall, 35 cases in the 20,000 cpm group and 37 cases in the 10,000 cpm group completed the trial without significant differences in baseline demographic characteristics. The 20,000 cpm group demonstrated significantly shorter core vitrectomy duration (161.6 ± 10.4 vs. 206.8 ± 10.1 s) (P = 0.003) and peripheral vitreous shaving time (446.3 ± 20.3 vs. 544.2 ± 22.2 s) (P = 0.002) compared to the 10,000 cpm group. BSS consumption was higher in the 20,000 cpm group but without statistical significance (P = 0.231). There were no significant differences in the need for scleral sutures and the incidence of iatrogenic retinal breaks (P = 0.331 and 0.523). At the 3-month follow-up, there were no statistically significant differences in primary success, final anatomical success, or mean visual acuity (P > 0.9, P = 0.326). Rates of complications, including ocular hypertension, epiretinal membrane formation, and endophthalmitis also showed no statistically significant differences (P > 0.6).

Conclusion

The utilization of the new-generation 27-gauge system with a 20,000 cpm probe may herald a new paradigm of high-flow, smaller-diameter instrumentation, thereby enhancing the efficiency of the small gauge technique.