Purpose <p>Cardiac magnetic resonance fingerprinting (cMRF) is a powerful quantitative imaging technique that provides multi-parametric diagnostic information. Here, we introduce an open-source framework for cardiac MRF including open-source pulse sequences, image reconstruction, and parameter estimation tools that are needed for the processing of the data.</p> Methods <p>A 2D cMRF sequence with a variable-density spiral readout is implemented using the open-source and vendor-agnostic sequence format Pulseq. Cardiac triggering is used to synchronize acquisition with the rest period of the heart. <InlineEquation ID="IEq1"> <InlineMediaObject> <ImageObject Color="BlackWhite" FileRef="10334_2025_1269_Article_IEq1.gif" Format="GIF" Height="16" Rendition="HTML" Resolution="72" Type="Linedraw" Width="17" /> </InlineMediaObject> <EquationSource Format="TEX">\(T_1\)</EquationSource> <EquationSource Format="MATHML"><math> <msub> <mi>T</mi> <mn>1</mn> </msub> </math></EquationSource> </InlineEquation> inversion and <InlineEquation ID="IEq2"> <InlineMediaObject> <ImageObject Color="BlackWhite" FileRef="10334_2025_1269_Article_IEq2.gif" Format="GIF" Height="16" Rendition="HTML" Resolution="72" Type="Linedraw" Width="17" /> </InlineMediaObject> <EquationSource Format="TEX">\(T_2\)</EquationSource> <EquationSource Format="MATHML"><math> <msub> <mi>T</mi> <mn>2</mn> </msub> </math></EquationSource> </InlineEquation> preparation pulses are added to ensure accurate parameter estimation. Data acquisition is carried out over 15 heartbeats. The images showing the signal changes over time are reconstructed and matched to a pre-calculated signal dictionary. In addition to the cMRF sequence, spin-echo reference sequences for quality control in phantoms are provided. The method is evaluated in phantom experiments using a T1MES phantom on four different scanners. In vivo experiments were performed to compare the open-source cMRF sequence with a vendor-specific cMRF sequence and clinical sequences used for <InlineEquation ID="IEq3"> <InlineMediaObject> <ImageObject Color="BlackWhite" FileRef="10334_2025_1269_Article_IEq1.gif" Format="GIF" Height="16" Rendition="HTML" Resolution="72" Type="Linedraw" Width="17" /> </InlineMediaObject> <EquationSource Format="TEX">\(T_1\)</EquationSource> <EquationSource Format="MATHML"><math> <msub> <mi>T</mi> <mn>1</mn> </msub> </math></EquationSource> </InlineEquation> and <InlineEquation ID="IEq4"> <InlineMediaObject> <ImageObject Color="BlackWhite" FileRef="10334_2025_1269_Article_IEq2.gif" Format="GIF" Height="16" Rendition="HTML" Resolution="72" Type="Linedraw" Width="17" /> </InlineMediaObject> <EquationSource Format="TEX">\(T_2\)</EquationSource> <EquationSource Format="MATHML"><math> <msub> <mi>T</mi> <mn>2</mn> </msub> </math></EquationSource> </InlineEquation> mapping of the heart. Three volunteers were imaged on two different scanners.</p> Results <p>The error of <InlineEquation ID="IEq5"> <InlineMediaObject> <ImageObject Color="BlackWhite" FileRef="10334_2025_1269_Article_IEq1.gif" Format="GIF" Height="16" Rendition="HTML" Resolution="72" Type="Linedraw" Width="17" /> </InlineMediaObject> <EquationSource Format="TEX">\(T_1\)</EquationSource> <EquationSource Format="MATHML"><math> <msub> <mi>T</mi> <mn>1</mn> </msub> </math></EquationSource> </InlineEquation> and <InlineEquation ID="IEq6"> <InlineMediaObject> <ImageObject Color="BlackWhite" FileRef="10334_2025_1269_Article_IEq2.gif" Format="GIF" Height="16" Rendition="HTML" Resolution="72" Type="Linedraw" Width="17" /> </InlineMediaObject> <EquationSource Format="TEX">\(T_2\)</EquationSource> <EquationSource Format="MATHML"><math> <msub> <mi>T</mi> <mn>2</mn> </msub> </math></EquationSource> </InlineEquation> over all tissue types present in the T1MES phantom was comparable between all four scanners and on average 4.50 ± 2.48%. <InlineEquation ID="IEq7"> <InlineMediaObject> <ImageObject Color="BlackWhite" FileRef="10334_2025_1269_Article_IEq1.gif" Format="GIF" Height="16" Rendition="HTML" Resolution="72" Type="Linedraw" Width="17" /> </InlineMediaObject> <EquationSource Format="TEX">\(T_1\)</EquationSource> <EquationSource Format="MATHML"><math> <msub> <mi>T</mi> <mn>1</mn> </msub> </math></EquationSource> </InlineEquation> and <InlineEquation ID="IEq8"> <InlineMediaObject> <ImageObject Color="BlackWhite" FileRef="10334_2025_1269_Article_IEq2.gif" Format="GIF" Height="16" Rendition="HTML" Resolution="72" Type="Linedraw" Width="17" /> </InlineMediaObject> <EquationSource Format="TEX">\(T_2\)</EquationSource> <EquationSource Format="MATHML"><math> <msub> <mi>T</mi> <mn>2</mn> </msub> </math></EquationSource> </InlineEquation> maps obtained in vivo were comparable between the open-source and vendor-specific implementation of cMRF.</p> Conclusion <p>The proposed open-source cMRF implementation enables accurate parameter estimation across multiple different scanners. Sequence files, image reconstruction, and parameter estimation scripts are available for reproducible quantitative MRI.</p>

错误:搜索内容不能为空,请输入英文关键词
错误:关键词超出字数限制,请精简
高级检索

Open-source cardiac magnetic resonance fingerprinting

  • Patrick Schuenke,
  • Catarina Redshaw Kranich,
  • Max Lutz,
  • Jakob Schattenfroh,
  • Matthias Anders,
  • Philine Reisdorf,
  • Jeanette Schulz-Menger,
  • Ingolf Sack,
  • Jesse Hamilton,
  • Nicole Seiberlich,
  • Christoph Kolbitsch

摘要

Purpose

Cardiac magnetic resonance fingerprinting (cMRF) is a powerful quantitative imaging technique that provides multi-parametric diagnostic information. Here, we introduce an open-source framework for cardiac MRF including open-source pulse sequences, image reconstruction, and parameter estimation tools that are needed for the processing of the data.

Methods

A 2D cMRF sequence with a variable-density spiral readout is implemented using the open-source and vendor-agnostic sequence format Pulseq. Cardiac triggering is used to synchronize acquisition with the rest period of the heart. \(T_1\) T 1 inversion and \(T_2\) T 2 preparation pulses are added to ensure accurate parameter estimation. Data acquisition is carried out over 15 heartbeats. The images showing the signal changes over time are reconstructed and matched to a pre-calculated signal dictionary. In addition to the cMRF sequence, spin-echo reference sequences for quality control in phantoms are provided. The method is evaluated in phantom experiments using a T1MES phantom on four different scanners. In vivo experiments were performed to compare the open-source cMRF sequence with a vendor-specific cMRF sequence and clinical sequences used for \(T_1\) T 1 and \(T_2\) T 2 mapping of the heart. Three volunteers were imaged on two different scanners.

Results

The error of \(T_1\) T 1 and \(T_2\) T 2 over all tissue types present in the T1MES phantom was comparable between all four scanners and on average 4.50 ± 2.48%. \(T_1\) T 1 and \(T_2\) T 2 maps obtained in vivo were comparable between the open-source and vendor-specific implementation of cMRF.

Conclusion

The proposed open-source cMRF implementation enables accurate parameter estimation across multiple different scanners. Sequence files, image reconstruction, and parameter estimation scripts are available for reproducible quantitative MRI.