Thermal insulation does not hamper assessment of injuries in trauma CT scans

  • Tomasz Sanak,
  • Aleksandra Skowronek,
  • Konrad Mendrala,
  • Tomasz Darocha,
  • Grzegorz Liszka,
  • Robert Chrzan,
  • Krzysztof Jerzy Woźniak,
  • Grzegorz Staskiewicz,
  • Paweł Podsiadło

摘要

Purpose

The use of thermal insulations reduces the risk of hypothermia, therefore decreases the risk of death in trauma victims. The aim of the study was to assess whether thermal insulations cause artifacts, which may hinder the diagnosis of injuries, and how the used thermo-systems alter the radiation dose in polytrauma computed tomography.

Methods

Computed tomography scans were made using the road accident victim body wrapped consecutively with 7 different covers. 14 injury areas were listed and evaluated by 22 radiologists. The radiation dose was measured using a dosimeter placed on the victim in the abdominal area.

Results

No significant artifacts in any of the tested covers were observed. The presence of few minor artifacts did not hinder the assessment of injuries. Certain materials increased (up to 19,1%) and some decreased (up to -30,3%) the absorbed radiation dose.

Conclusions

Thermal insulation systems tested in this study do not cause significant artifacts hindering assessment of injuries in CT scans. Concern for artifacts and increased radiation dose should not be a reason to remove patients’ thermal insulation during performing trauma CT-scanning.