Electrochemical behavior of Co(II) polymer-modified glassy carbon electrode for the determination of dopamine
摘要
A new electrochemical sensor for the highly sensitive and selective determination of dopamine was developed, based on the use of a glassy carbon electrode modified with cobalt(II) polymer. Both electrochemical and morphological characterizations were investigated. Square wave voltammetry (SWV) was used for the determination of dopamine. From the SWV results obtained on the modified electrode under the optimized conditions, the successive addition of dopamine increased the anodic peak current. The results obtained show a linear response of the electrode modified with Co(II) polymer in a wide concentration range. Each concentration decade is characterized by a new calibration plot with a corresponding mathematical expression. The highest sensitivity (13.182) was obtained for the lowest concentration range, and the detection limit of 0.03 µM, calculated based on the 3 × sb/m criterion, indicates a very high sensitivity and great application potential. When analyzing a real sample, the modified electrode showed good electroanalytical activities in terms of oxidation of dopamine, good anti-interference ability, stability, reproducibility, and potential applicability in commercial pharmaceutical samples.