Intraoral vs. extraoral bitewing radiography for approximal caries detection: A multi-observer ex vivo ROC study using thin-section microscopy as gold standard
摘要
This ex vivo study aimed to compare the accuracy in detection of interproximal natural carious lesions between intraoral (iBWR) and extraoral bitewing radiographs (eBWR) using a multi-observer design and a rigorous gold standard.
Materials and methodsEighty extracted teeth (40 premolars, 40 molars) were arranged in anatomical sequence within a simulated jaw composed of PMMA and modified gypsum, with an emphasis on creating natural interproximal contacts. Approximately 50% of the teeth exhibited enamel caries, while the remaining 50% were caries-free. Image acquisition was performed using a custom-designed PMMA phantom. iBWR were obtained with a CMOS intraoral sensor (XIOS XG Supreme, Sirona Dental Systems, Bensheim, Germany), and eBWR with a digital panoramic device (Orthophos SL 3D, Dentsply Sirona, Bensheim, Germany).
Twenty-seven licensed dentists assessed caries presence and depth on 120 approximal surfaces (each surface assessed twice using both modalities) using a 5-point confidence scale and a 4-point lesion depth scale. Observers were blinded to the true caries status, which was determined through histological serial sectioning and brightfield microscopy. Diagnostic accuracy was evaluated via ROC analysis, with Youden’s index used to calculate sensitivity, specificity, predictive values, and likelihood ratios. Statistical analyses were conducted at a significance level of α = 0.05.
ResultsOverall accuracy was higher for iBWR (Azpooled = 0.58) than for eBWR (Azpooled = 0.54). Both intra-rater (test-retest, eBWR
Overall, our multi-observer ex vivo study using microscopy as ground truth revealed higher diagnostic accuracy for intraoral bitewing radiography as compared to its extraoral counterpart.
Clinical RelevanceOur results from a highly standardized study using a rigorous gold standard support the assumption that intraoral bitewing radiography still represents the radiographic state-of-the-art in interproximal caries detection. For minute enamel, diagnostic accuracy of both methods is just above random guessing.