Impact of enhancement filters of a CMOS system on halo artifact expression at the bone-to-implant interface

  • Manuella Soussa Braga,
  • Ana Maria de Almeida Ramos,
  • Fernanda Coelho-Silva,
  • Eduarda Alberti Bonadiman,
  • Teresa Cristina Rangel Pereira,
  • Sergio Lins de-Azevedo-Vaz

摘要

Objective

To assess the impact of enhancement filters on the formation of halo artifacts in radiographs of dental implants obtained with a complementary metal oxide semiconductor (CMOS) system.

Methods

Digital radiographs of dental implants placed in dry human mandibles were processed with the Noise Reduction smoothing filter, as well as the Sharpen 1, Sharpen 4, and Sharpen UM high-pass filters available in the CLINIVIEW™ software (Instrumentarium Dental, Tuusula, Finland). Subjective analysis involved evaluating the left, right, and apical surfaces of each implant for the presence of much, few, or no halo. The objective analysis involved measurement of the halo area using the Trainable Weka Segmentation plugin (ImageJ, National Institutes of Health, Bethesda, MD, USA). Data were analyzed using Friedman’s test (subjective analysis) and ANOVA (objective analysis) (α = 5%).

Results

In the subjective evaluation, the Sharpen 4 filter produced more radiographs with much halo present, and in the objective evaluation, a bigger halo area when compared to the original images and the Noise Reduction filter for all surfaces (p < 0.05).

Conclusions

When evaluating dental implants, priority should be given to original images and those enhanced with smoothing filters since they exhibit fewer halo artifacts.

Clinical relevance

Post-processing tools, such as enhancement filters, may improve the image quality and assist some diagnostic tasks. However, little is known regarding the impact of enhancement filters in halo formation on CMOS systems, which have been increasingly used in dental offices.